|
First published in
EOS/ESD Technology Feb/March 1989
Measure Resistance Instead: Part
II
Resistivity and decay measurements
can not only be misleading,
but can also be sometimes confused by misunderstandings
and misused terminology. The simple, unambiguous test described
here can provide much more reliable materials characterizations.
Link to
Measure Resistance Instead: Part I
Ben Baumgartner
Lockheed Missiles & Space Co. Inc.
Sunnyvale, CA
Part 1 of this article
(EOS/ESD Technology, Dec./Jan. 1989, p. 20) discussed some of the
technical problems of materials characterization using resistivity
measurements. Part 2 continues this discussion and proposes a simple,
almost foolproof test for characterizing ESD-protective materials
and packaging in the real world.
Circuit Models
of Materials.
Considering only the
resistive component, a circuit of an ideal material would look like
the example in Fig 1a. The surface cell resistivity, Rs,
for this ideal material is much less than the volume-cell resistivity,
Rv. A surface-resistivity measurement would show RS cells
as 40 W (adding up surface cells) without a large error because
RS is little affected by the shunt effect of RV
 |
| Figure 1. Models of insulative
(a) and buried-layer (b) materials. |
The horizontal volume
cells add up to 400 W and a path from the top to the bottom surface
along the bottom and back to the top would add up to 120 W (four
volume units X 10 W + 40 surface units X 1W + four volume units
X 10 W). The current through this resistance path to te guard electrode
is subtracted from the surface-resistivity measurement. When the
value of RV becomes lower than ideal (insulative), it will affect
the surface-resistance measurement. If the resistivity of the volume
cells is changed to 1 W, the volume cell can be seen to dominate
the measurement.
In terms of the number
of cells, the length-to-thickness ratio of volume cells is 10 to
1 and wold translate into a 0.040-in. electrode
spacing for a 0.004-in. (4-mil) material. ASTM D-257 requires a
gap (g) of at least twice the material thickness, and in most circular-electrode
designs this ratio is much larger when measuring 4-mil material.
If a buried layer of
lower resistivity, or conductive material, is added, the example
circuit could look like Fig. 1b. The resistivity value of RV per
cell would not change, but the dominant path would be drastically
altered.
The path from the surface
of the material to the buried layer is short (two cells- 20 ohms),
so the resistance is very low to conductive layer C. The predominant
path has changed the point-to-point surface resistance to 40 W shunted
by 44 W (2 V X 10 + 40 cell X 0.1 + 2 V X 10). The measurement is
no mostly a volume measurement and is almost independent of electrode
spacing.
But this is only an
illustration; in any real material, resistance is more complex.
For example, the surface-resistivity concept becomes extremely difficult
to understand if applied to tabletop and garment materials.
Volume measurements
encounter similar difficulties when different resistivity layers
are present in a material. The layer with the highest resistivity
dominates the measurement, and yet the thickness of all the layers
will be used in calculating the resistivity. Wires or other conductive
materials that may or may not be confined to a layer can result
in measurements that simply can't be related to the concept of volume
resistivity.
|
Resistance,
Protection and Circuit Packaging
|
The reason for measuring package
materials can be seen by following the two possible current
paths in Fig. 1. (The three figures here show the equivalent
circuits of ESD paths.)
One possible discharge path is through the package wall, after
which the IC pin carries the current (which goes through the
component) to a pin on the opposite side of the circuit board.
Finally, the current passes through the package to ground.
The magnitude of this current must be considered for two types
of sensitive devices. Diode junction or bipolar devices are
protected if junction currents are limited to a dissipation
rate that the devices can handle. If these currents are small,
the voltage drops in the devices will not produce arcs across
metallizaed IC traces, nor will these small currents melt IC
traces or silicon in the junction areas.
These type of devices are energy sensitive. The package resistance
can be specified and provides protection from low-frequency
current.
The metal-oxide semiconductor in Fig 2 is voltage sensitive.
A voltage-sensitive device passes no current thorough the oxide
at voltages below oxide-breakdown potential. Even if a large
amount of energy is available from a source, no energy is dissipated
in the oxide below the breakdown voltage. However, when oxide
breakdown voltage is exceeded, the device can be damaged by
an exceedingly small amount of energy.
Because voltage breakdown is necessary for small currents to
flow, unprotected oxide is considered voltage sensitive. (When
an oxide device is protected by a diode, the device acts like
an energy-sensitive device and can be classed as such).
For voltage-sensitive devices, resistances have been measured
as high as 1 X 1014, including package-header resistance
(which may account for much of what is measured). This high
resistance causes extremely small currents passing through the
bag to generate an voltage drop across the gate oxide. Therefore,
currents caused by tribocharging the top surface could damage
high-resistance, voltage-sensitive devices. Proper volume resistance
measurements can determine the package resistance required for
a given surface potential to limit the voltage drop across such
a high device resistance.
A drop of half the source voltage would require a total package
resistance equal to the resistance of the device being protected.
This is not sufficient to protect voltage-sensitive devices
without the package providing an external shunt resistance to
drop the source voltage to a manageable level. The resistance
paths through the material and the conductive (metallized) layer
shunt resistance, 2Rs (both sides), including the seam resistance,
are shown in Fig 3.
Package protection can only be estimated using resistances because
the high-frequency effects of inductance, Ls, and
seam bond capacitance, Cb, are not considered. |
 |
 |
 |
| Figure
1. A pictorial of circuit paths in an ESD bag. |
Figure
2. The path through an ESD bag and a sensitive IC. |
Figure
3. Equivalent circuit in a metallized layer bag. |
|
Problems with
Methods
The point-to-point
resistive components of a material are not given by Method 4046,
and are not easily determined by the method in ASTM D-257 due to
interrelated problems with materials and methods. An illustration
of the problems was seen in Fig 4 and 5 of Part 1 for a buried-layer
material, and applies to both measurement methods. Neither instrumentation
method can separate surface and volume current for all ESD-protective
materials.
If surface resistivity
is measured without a guard electrode, the measurement is essentially
a volume measurement, as discussed earlier, but, frequently, unguarded
measurements occur when the electrodes cannot conform to irregular
or molded materials, and, therefore, make poor contact.
There are a large variety
of unguarded electrode sizes and designs that also affect the volume-current
component of such a measurement. Electrode-contact areas change
with the use of knife edges, braid or rubber electrodes of various
configuration and should be expected to cause differences in measured
values, as shown in the table. This table's measurements were made
on tabletop materials with five different electrode designs.
|
RESISTANCE
AND RESISTIVITY MEASUREMENT
|
|
[ raw data/
(converted data) ]
|
|
Probe Number
|
|
Mat
|
1
|
2
|
3
|
4
|
5
|
|
1
|
1 E8 |
(2E9) |
2E10 |
(4E11) |
5E10 |
2E8 |
2E8 |
|
2
|
1 E8 |
(2E9) |
1.5E10 |
(3E11) |
9E10 |
1.5E8 |
1.5E8 |
|
3
|
6 E8 |
(1.1E9) |
9E8 |
(1.8E9) |
2E9 |
1.5E9 |
4E9 |
|
4
|
1 E10 |
(2E11) |
2E12 |
(2E13) |
6E12 |
2E11 |
5E10 |
| Note: Single values
are both raw and converted data. |
[EOS/ESD Technology
Editor's note: Although tabletop materials were measured as
an example, this method is not proposed for tabletop measurements.
Though similar, this method is different from the EOS/ESD standard
worksurfaces measurements in that resistance is measured rather
than surface resistivity. Baumgartner supports the measurement
methods used in the EOS/ESD Association's worksurface standard,
and is a member of the worksurfaces committee. The two measurement
methods should not be confused.]
For volume measurements,
multiple possible current paths can prevent accurate measurements.
The presence or absence of a guard is an unknown variable, unless
described in a material's data sheet, and affects the surface currents,
Is, present in the measurement. Surface current may also
be present in volume measurements made on small samples-- samples
that will usually differ from large flat specimens of the same material.
Volume current will
be increased by a buried conductive layer or composite cloth with
conductive-thread groups. The lower electrode in Fig 5 of Part 1
could receive increased current due to the material's conductive
buried layer. This is significant if the E1 area is used for calculating
volume resistivity, but the true area actually being measured is
the unknown area contacting the lower electrode.
There may be as many
electrode-configuration interactions with ESD materials as there
are material types. Certainly no user can be expected to research
or understand all of these problems. So resistivity measurements,
while proper for ideal materials, are usually inappropriate for
characterizing ESD materials in actual use.
In fairness, surface-
and volume- resistivity measurement problems are not due simply
to the instrumentation involved. When used properly on the right
material, good instrumentation will produce correct values.
A Simple, Two-Point
Resistance Measurement
Although resistivity
measurements are useful during development of barrier materials
and design of packaging systems, conditions outside the laboratory
lead to the sort of measurement inaccuracy and confusion just described.
However, ESD-protective bags, totes, and boxes need to be tested
in the workplace, both before and after use. Blister packs and other
specially formed packages must also be tested nondestructively.
Obviously, a test method
suited to real-world use is needed. Therefore, we propose a simple,
two-point measurement of resistance for use with dissipative-range
materials.
The method uses 0.125-in-dia
electrodes 0.75 in. apart for surface-type measurements. Surface
resistance is measured with the electrodes on the same side of the
material, while volume-mode measurements are made with electrodes
on opposite sides of the material, and in line with each other.
Using this approach,
any finished material can be measured in a specific location. Materials
that have visible cells or pores, such as foams, can be characterized
as well as more homogeneous samples, such as sheets, molded sheets,
laminates, composites, cloths, and bubble packs. A measurement protocol
would have to be established for some composite materials.
|
|
|
| Figure 2. Diagrams
of a two-point surface measurement (a) and a two-point volume
measurement (b). |
Fig 2a shows the equivalent
circuits for surface and volume measurements. Any size, shape, or
construction of ESD-protective dissipative material can be measured
with this approach, including foams, composites, laminates, dissipative
cloth, etc.
The two-point, surface-resistance
(RS) measurement shown in Fig 2b produces consistent measurements.
Interelectrode distance is fixed for all surface measurements so
that consistent readings result, regardless of the presence or absence
of a metallized resistive film RM. Also, electrode voltage
is fixed to produce consistent values since resistance values may
be nonlinear with voltage.
Electrode spacing for
surface-mode measurements is 0.75 in; 100 V is the applied open-circuit
electrode voltage. The same voltage is used for volume-mode measurement.
This voltage was selected because 100 V is often the specified maximum
allowable potential between surfaces in ESD-safe work areas.
In addition, 100 V
is a reasonable voltage should a material's resistance become nonlinear
below this level, halting current flow. Based on 100 V and 0.75-in.
Electrode-spacing values, the linear IR drop of points spaced 1
ft. apart would be equivalent to 1600 V (100 V/0.75 in. X 12 in.).
Buried-layer materials are nonlinear and, therefore, have a lower
equivalent voltage for a 1-ft separation depending on the volume
conductivity to the metallized layer.
This method involves
the use of a 100-V power supply to place a voltage on the material
or the exposed electrodes. To avoid shock, the supply must be limited
to safe currents preferably below the shock-sensation level. If
this is done, the equipment may be used like any other low-voltage
instrument. A 100 V source with current limited to 0.1 mA would
be able to measure down to a resistance of 1 MW.
Significance and
Use
The purpose of this
procedure is to evaluate the surface and volume resistance of materials
used to protect ESD-sensitive devices. Surface-mode resistance is
used to characterize the ability of a material to dissipate static
charge from its surface, based on a fixed electrode size and spacing.
Electrodes for surface and volume measurements are shown in Fig.
3.
|
|
| Figure 3. Surface and volume
resistance of materials used to protect ESD-sensitive devices
are measured via a two-point electrode method. |
Volume-mode resistance
characterizes the ability of a material to dissipate static charge
throughout itself, based on a fixed-electrode-size measurement for
the actual material's thickness. Volume resistance is particularly
important in evaluating packaging for voltage-sensitive devices.
Because resistance, and not resistivity, is being measured, this
method has none of the material restrictions required of ASTM D-257
resistivity measurements. In a straightforward resistance measurement,
it is clear that both surface and volume currents are involved.
This method evaluates
the current path resistance for dissipative materials of all forms
and types and accurately
simulates worst-case conditions.
When volume resistance
is measured using this procedure, a portion of the measurement current
may flow on the surface of the materiel. Likewise, when a surface
charge is dissipated from a material, a portion of it may flow on
the material's surface. Therefore, even though the resistance measured
by this procedure includes both surface and volume resistance, the
current paths measured are a closer simulation of the actual path
through which a charge is dissipated.
Extremely accurate
measurements of volume or surface resistance for ESD-protective
materials are not required because actual discharge paths are uncertain.
However, measurements to within an order of magnitude are possible--
an improvement over the present methods using a variety of electrode
configurations.
The resistance values
yielded by this method and resistivity as measured by ASTM D-257
are not comparable. Although both this method and ASTM D-257 produce
about the same numerical values for ideal surface measurements,
the values differ more for volume measurements because this method
does not normalize the measurement for the thickness of the specimen
involved. Also, no guard electrode is involved in the measurement,
so unrestricted current paths can be similar to the unguarded static-charge
paths in a finished package.
The values given by
this method are not intended to classify material as dissipative.
However, if desired, finished products can be determined to be in
the dissipative range using this method's measured values rather
than those determined by ASTM D-257.
Note that simply measuring
a resistance value in the dissipatvie range does not necessarily
qualify a material for an intended application. On the other hand,
a value as measured by this procedure can be used to provide a more
desirable finished product instead of just meeting the blank criterion
of being in the dissipative range.
Comparing resistance
measurements for two products of vastly different thickness prodeces
a comparison of volume-mode charge-dissipation time. This can be
accomplished only with a resistance measurement of an actual material
thickness, and not with normalized ohm-cm measurements.
For data sheets describing
materials, resistance can be reported for a 3- X 3- in. (7.62- X
7.62-cm), flat standard specimen. ESD-protective products (finished
goods) may be tested "as is." All these items, when measured
as lab qualification specimens, must be conditioned at a stated
humidity.
|
EOS/ESD
101: Resistance versus Resistivity
|
|
In measuring the electrical properties of
materials, the terms resistance and resistivity both are used.
They are not the same. Here's how to tell the players without
a scorecard.
When voltage is applied to a circuit of conductive material,
current flow in that material encounters difficulty in moving
from one point to another through the material. This difficulty
can be measured, and is called resistance. This resistance
is a function of the material's resistivity as well as of
other physical parameters, including the dimensions of the
conducting path.
Since copper wire has very low resistivity,
a foot of it offers relatively little resistance to the flow
of current. However, if all other parameters are kept equal,
but the length of the wire is increased to a mile, resistance
will rise more than 5,000 fold. On the other hand, the resistivity
of a pure sample of a given element is constant and is independent
of the material's dimensions.
Both conductors and insulators have resistivity. If we were
to use a foot of copper wire in an electrical circuit, the
twine "circuit" would not work. The twine has such
a high resistivity that the resulting resistance of the circuit
would make current flow next to impossible.
Why make an issue of this? Have you ever supplied a sample
of a material or a new bag to an engineer, then watched him
use an ohmmeter to measure its "resisitivity" and
then had to listen as he complained that the material's resistivity
spec was wrong? This happens too often, and, of course, it's
because the engineer is really measuring the resistance between
the test probes, not the resistivity of the material.
The two-point test proposed in Ben Baumgartner's article could
overcome this problem by making the obvious resistance measurement
the correct measurement as the user level. For designers of
packaging and barrier materials, resistivity would still be
an appropriate and useful specification, but life would be
a lot simpler for most of us.
But not oversimplified. Baumgartner's two-point method specifies
a manner in which such resistance measurements are to be made
because a variety of peripheral issues can affect the accuracy
of these measurements to varying degrees.
Some of these sources of error (e.g., contact resistance,
sample resiliency, etc.) will be important only to those who
need maximum precision, but that's not the point. They all
count, and the sort of sloppy thinking that confuses resistance
and resistivity can also lead to sloppy measurement.
Charles R. Hynes, Atrix, Inc., Contributing Editor.
|
Summary and Conclusions
Specifications and
measurement of dissipative materials must give an accurate representation
of products used to protect sensitive devices. Unfortunately, this
is not always true.
Resistivity measurements
(ASTM D-257) are useful for engineering data, but inappropriate
for characterizing ESD materials in use. The static-decay test (FED-STD-101C,
Meth. 4046) lacks the ability to cover the dissipative range, and
can thus give a confused impression of surface-resistivity.
Circuit models explain
some of the variations in resistivity and dissipation-time measurements,
and the cause of these variations is a combination of instrumentation
and material interactions. The two-point resistance measurements
described here can reduce the confusion and improve the intended
package protection.
Protective designs
for energy-sensitive and voltage-sensitive devices must be assessed
differently, and require realistic measurements. This test method
determines the surface and volume resistance of static-dissipative
materials by measuring current when a voltage is applied on the
surface, or across the volume, of a dissipative material.
References
1. D.C. Burdeaux, C.L.
Mott, "An Analytical Approach to Surface Resistivity Measurement,"
Evaluation Engineering, November 1986 (Table 3, p. 87; Summary,
p. 97).
2. B.N. Stevens, "Determining the Surface Resistivity of ESD-Protective
Cellular Packaging Materials," Proceedings, 1986 EOS/ESD Symposium.
3. S.L. Fowler, "Triboelectricity and Surface Resistivity Do
Not Correlate," Proceedings, 1988 EOS/ESD Symposium.
4. Packaging Material Standards for ESD-Sensitive Items,
Electronic Industries Assn., EIA STD IS-5-A (now EIA-541; 2.2.2,
4.2.3).
5. V.E. Shashous, "Static Electricity in Polymers-- Theory
and Measurement," Journal of Polymer Science,"
1958, 32:65-85.
6. G. Baumgartner, R. Havermann, "Testing of Electrostatic
Materials-- Fed. STD 101C, Method 4046.1," Proceedings, 1984
EOS/ESD Symposium.
7. G. Baumgartner, "Static Decay Test versus Resistivity Measurements,"
EMC EXPO, 1987 International Conference on Electromagnetic Compatibility.
This article (Parts
I and II) is based on a paper that was originally presented at the
1987 EOS/ESD Symposium and is reproduced here in revised and updated
form with the permission of the EOS/ESD Association.
|